Features ;
> 3x0.2mL reaction module.
> Portable small size and lightweight design for outdoor experiments.
> 4-color fluorescence channels applicable to most tests.
> Easy-to-use bluetooth tablet connection
Application Area
> Basic Research
In research on molecular cloning, gene expression and genotyping in scientific research institutes,
> Medical Diagnosis
Pathogen detection, genetic screening, tumor diagnosis, etc. in medical institutions.
> Food Safety
Entry-Exit and FDA inspections of foodborne pathogens, genetically modified products, meat adulteration, excessive microorganisms, etc.
> Epidemic Prevention and Control
Identification and tracking of epidemics (avian flu, SARS, dengue fever, etc.) by national systems such as CDC and Customs.
Technological Innovation
> Innovation in Temperature Control
Based on six-zone independent temperature control, using the high efficiency of closed liquid conduction, combines with environmental scanning monitoring and auxiliary infrared heating adjusted by intelligent frequency conversion of voltage and current system to ensure a fast ramp speed and effectively prevent Temperature overshoot without edge effect and evaporation results in better temperature uniformity and repeatability in amplification, saving time and increasing efficiency.
> Optical Detection Advantages
Representing the world's leading optical transmission and collection technology, optical fiber is used to transmit excitation and emission light to each reaction well and cooled CCD without attenuation.It improves detection sensitivity to ensure effective detection of low-copy samples while ensuring consistency and authenticity of excitation and detection.The added channel mismatch functionality extends the scope of application of qPCR to the protein level and provides a new way to generate multiple systems of diagnostic reagents.
> Improved Software
Real-time monitoring, automatic identification and calculation of positive and negative results, automatic generation of standard curve, absolute/relative quantification, multiplex quantification, melting curve, gene mutation, quality control chart analysis, fluorescence calibration, PCR amplification efficiency, etc.
Technical Specifications ;
Basic |
|
Sample Capacity |
3x0.2mL |
Applicable Consumable |
0.2mL Single Tube |
PCR Volume Range |
20-120uL |
Working Temperature |
15-35°C |
Storage Temperature |
-20-55°C |
Ambient Relative Humidity |
<85% |
Dimensions and Weight |
130x130x220mm (WxDxH), 2 Kg |
Power Supply |
100/240V, 50/60Hz, 180VA (optional external battery) |
Temperature Control System |
|
Heating/Cooling Method |
Peltier, independent and intelligent temperature control |
Temperature Range |
4°C-100°C |
Ramp Rate |
Heating : 7.5°C, Cooling : 7.5°C |
Temperature Accuracy |
±0.1°C |
Temperature Uniformity |
±0.15°C |
Detection System |
|
Excitation Source |
LED |
Detection Device |
PD |
Light Propagation Medium |
Optical fibers for space |
Detection Channels |
4 |
Excitation Range |
1st Channel : 470nm±10nm 2nd Channel : 525nm±10nm 3rd Channel : 570nm±10nm 4th Channel : 628nm±10nm |
Detection Range |
1st Channel : 520nm±10nm 2nd Channel : 570nm±10nm 3rd Channel : 628nm±10nm 4th Channel : 670nm±10nm |
Dye Compatibility |
FAM/SYBR Green/Eva Green/LC Green/Fluorescein; VIC/HEX/TET/Cy3/JOE/Alexa555; ROX/Cy3.5/Texas Red; Cy5/LC Red640 etc. |
Detection Sensitivity |
>1 copy |
Confidence Coefficient |
99.90% |
Sensitivity |
Allowed 1.5-fold in single reaction |
Sample Detection Repeatability |
CV<1% |
Linear Range |
1-1010 |
Sample Linearity |
>0.99 |
Software |
|
Software Language |
English |
Control Method |
Bluetooth tablet connection, breakpoint signal acquisition |
Software Function |
Real-time monitoring, automatic identification and calculation of positive and negative results report export. |
Output |
Excel/Word/PDF |